power quality analyzer

  • It can measure the total harmonic distortion rate of voltage and current
    It can measure the 0.5-31.5 Inter-harmonic content of voltage and current
Contact Us+86 13720145047

Details

 Functions and features

1. Multi-channel testing: Simultaneous measurement of 4 voltage channels and 4 current channels.

2. Electrical parameter measurement: It can measure a variety of electrical parameters at the same time, such as voltage, current, active power, reactive power, phase angle, power factor, frequency, etc.

3. It can measure the voltage harmonic and 2-64th current harmonic content

4. It can measure the 0.5-31.5  Inter-harmonic content of voltage and current

5. It can measure the total harmonic distortion rate of voltage and current

6. Can measure and short-term flicker (Pst), long-term flicker (Plt), voltage fluctuation

7. It can measure positive sequence voltage, negative sequence voltage, zero sequence voltage, voltage unbalance

8. It can measure positive sequence current, negative sequence current, zero sequence current, current unbalance

Technical specifications






Project

Data


Number of measurement channels

Four voltage channels, four current channels


Measuring range

Voltage

0-660V


Current

Small clamp:diameter 8mm,5A/25A(standard)

Medium clamp:diameter 50mm,100A/500A(Optional)

Large clamp:diameter 125×50mm,400A/2000A(Optional)

Rogowski Coil : Ring diameter 250mm, 1000A/5000A(Optional)


Phase angle

0.000-359.999°


Frequency

42.5-69Hz


Measuring step

Voltage

0.001V


Current

0.0001A


Phase angle

0.001°


Power

active power 0.01W、reactive power 0.01Var


Frequency

0.0001Hz


Voltage measurement accuracy

≤0.1%


Current measurement accuracy

≤0.3%


Phase angle measurement accuracy

≤0.


Power measurement accuracy

≤0.5%


Frequency measurement accuracy

≤0.01Hz


Harmonic analysis times

2-64 times


Harmonic voltage measurement error

Harmonic content>1% Nominal value≤1% RD

Harmonic content1% Nominal value≤0.05%FS


Harmonic current measurement error

Harmonic content>3% Nominal value≤1% RD

Harmonic content<3% Nominal value≤0.05%FS


Three phase voltage unbalance error

0.2%


Three phase current unbalance error

0.5%


Voltage fluctuation accuracy

5%


PST measurement time

10min


PLT measurement time

2h


PST/PLT measurement error

5%


Screen

128800True color LCD


Charging Power

AC220V±15%

Frequency 45Hz-65Hz


Standby timing

10h


Host Power

4VA


Insulation

1) The insulation resistance between the voltage, current input terminal and the housing 100MΩ.

2) Withstand 1.5KV (valid value) with power frequency on operating power input and the shell, which lasts one minute.


Working temperature

-2050


Relative humidity

0-95% No condensation


Size

280mm×210mm×58mm(L×W×H)


Weight

2kg




Hotline

Service hotline

+86 13720145047