Product Parameter
Specification  | 120/2  | 120/3  | 120/5  | 
Rated(kV)  | 120  | 120  | 120  | 
Rated(mA)  | 2  | 3  | 5  | 
Rated power(W)  | 240  | 360  | 600  | 
Weight(kg)  | 10  | ||
Volume(mm3)  | 565*390*190  | ||
Voltage cylinder height(mm)  | 535  | ||
Accuracy of output voltage  | ±(1.0%R±2D)  | ||
Accuracy of output current  | ±(1.0%R±2D)  | ||
Ripple factor  | ≤0.5%  | ||
Working mode  | Intermittent use, 30 minutes under-rated load  | ||
Over-load capacity  | The no-load voltage can exceed the rated voltage of 10% for 10 minutes  | ||
The maximum charge current is 1.25 times the rated current  | |||
Power  | AC220V±10% 50HZ  | ||
Service conditions  | Temperature: -10-40℃  | ||
Relative humidity: less than 85% under 25℃, without moisture  | |||
Altitude:1500M under  | |||
Product Feature And Application
1.Output voltage stable: Adopts the latest PWM pulse width modulation technology and the voltage and current double closed-loop feedback technology. The power supply adjustment rate and load regulation rate is improved, with a small ripple.
2.Comprehensive protection: Completed protection functions, like zero protection, over-voltage protection, over-current protection, breakdown protection, and circuit protection, using nanosecond special sensor, fast and reliable action, can effectively protect user and equipment safety.
3.0.75U function: add intelligent high-precision 0.75U function button. With this button, the voltage and current automatically to the 0.75U state, which is conducive to the zinc oxide arrester test.
4.Voltage stepping from zero: boost potentiometer zero boosts, with multi circle potentiometer, boosting process is stable, high precision adjustment.
5.Over-voltage setting: use a digital dial switch, easy to operate, high setting accuracy.
6.Integral design: The cylinder and the host are placed in a box. Use imported high-frequency high-voltage rectifier diodes, the cylinder is compact in size, portable, and easy to carry.
7.Reliable performance: the key components are high-performance imported components, the outer surface of the cylinder is coated with special insulation materials and has good electrical performance, strong moisture capacity, and no leakage.
8.The operation is simple: the instrument interfaces each function key, the layout is reasonable, the instruction is clear, and easy to learn, and use
Accessory
Number  | Name  | Amount  | Number  | Name  | Amount  | 
1  | Mianframe  | one  | 8  | Four-core connecting line  | one  | 
2  | High voltage cylinder  | one  | 9  | Ground lead  | one  | 
3  | Micro-ampere meter  | one  | 10  | 3A Fuse  | two  | 
4  | Discharging rod  | one  | 11  | Specification  | one  | 
5  | Current limiting resistor  | one  | 12  | Inspection Report  | one  | 
6  | High Voltage power line  | one  | 13  | Certification  | one  | 
7  | Power line  | one  |