Product Parameter
1 | Use conditions | -15℃∽40℃ | RH<80% | ||
2 | Anti-interference | Frequency conversion | |||
3 | Power | AC 220V±10%/110V±10%,50/60HZ | Allowable generator | ||
4 | High output | 0.5KV∽10KV | Every 0.1kV | ||
Precision | 2% | ||||
Maximum current | 200mA | ||||
Capacity | 2000VA | ||||
5 | Self-excited power | AC 0V∽50V/15A | 50HZ 60HZ 45HZ/55HZ 47.5HZ/52.5HZ 55HZ/65HZ 57.5HZ/62.5HZ Automatic double frequency | ||
6 | Resolving power | tgδ: 0.001% | Cx: 0.001pF | ||
7 | Precision | △tgδ:±(Reading*1.0%+0.040%) | |||
△C x :±(Reading*1.0%+1.00PF) | |||||
8 | measuring range | tgδ | Unlimited | ||
C x | 15pF < Cx < 300nF | ||||
10KV | Cx < 60 nF | ||||
5KV | Cx < 150 nF | ||||
1KV | Cx < 300 nF | ||||
CVT test | Cx < 300 nF | ||||
9 | LCR range | L>20H(2kV) | R>10KΩ(2kV) | ||
LCR Precision | 1% | Angular resolution | 0.01 | ||
10 | CVT range | 10∽10000 | |||
accuracy | 1% | ||||
Resolving power | 0.01 | ||||
11 | Exterior dimension | 350(L)×270(W)×315(H) | |||
Exterior dimension | 350(L)×270(W)×160(H) | ||||
11 | Memory size | The 200 group supports U disk data storage | |||
12 | Weight | 22.75+5.25(kg) |
Product Feature And Application
1.Large LCD display
Simple operation, the instrument is equipped with a high-end full touch LCD screen, and super touch interface, each process is very clear, and the operator does not need additional professional training can be used. Click on it to complete the whole process of measurement, is a very ideal intelligent dielectric loss measurement equipment.
2.Mass storage data
The instrument is equipped with a calendar chip and large capacity memory, the test results can be stored in chronological order, at any time to view the history, and can print output.
3.Multiple test modes
The instrument can use high pressure, high voltage, standard, standard, positive connection method and reverse connection method, self- excitation method and other methods can do testing; high voltage in standard high voltage conditions (greater than 10kV) dielectric loss.
4.CVT test step in place
The instrument can also test the fully sealed CVT (capacitor voltage transformer) C1, C2 dielectric loss and capacitance, to achieve the C1, C2 simultaneous test. The instrument can also test the CVT ratio and voltage angle difference.
5.Do not dismantle the high-voltage lead measurement CVT
The instrument can accurately measure the dielectric loss and capacitance of CVT without dismantling the CVT high voltage lead.
Accessory
SN | Designation | QTY. |
1 | Host computer | 1 unit |
2 | Accessory case | 1 pc |
3 | Red HV test wire | 1 pc |
4 | Black LV test wire | 1 pc |
5 | AC220V power cord | 1 pc |
6 | CVT self-excited power cord (attached) | 1 pc |
7 | Ground wire | 1 pc |
8 | Operation Instruction | 1 copy |
9 | Certificate of Conformity | 1 copy |